Welcome to this issue of touchREVIEWS in Neurology, where we explore significant advances in neurology, cognitive health, and wearable technology in the management of various chronic conditions. This issue brings together a collection of expert perspectives and research that spans innovative therapies, preventive strategies, and case studies, each offering critical insights for clinicians and researchers. […]

7 mins

Are All Gadolinium-based Contrast Agents Similar? The Importance of High Stability, High Relaxivity and High Concentration

Nicoletta Anzalone
Share
Facebook
X (formerly Twitter)
LinkedIn
Via Email
Mark CompleteCompleted
BookmarkBookmarked
Copy LinkLink Copied
Download as PDF
Published Online: Jun 4th 2011 European Neurological Review, 2009;4(2):98-102 DOI: http://doi.org/10.17925/ENR.2009.04.02.98
Select a Section…
1

Abstract

Overview

Gadolinium-based contrast agents (GBCAs) are commonly used to enhance image acquisition via magnetic resonance imaging, but they differ in their physicochemical characteristics and therefore their function. The stability, concentration and relaxivity of a GBCA can have a major impact on clinical efficacy. Stability is related to safety. GBCAs can be categorised into three stability classes: non-ionic linear agents, ionic linear agents and macrocyclic agents, in order of increasing stability. Relaxivity and concentration are contributing factors to the level of enhancement that can be achieved with any given GBCA. Gadobutrol, gadobenate and gadoxetic acid have relatively high relaxivities. Among these compounds, only gadobutrol is commercially available at a higher concentration of 1.0M, whereas the others are used at a standard concentration of 0.5M. With high stability, relaxivity and concentration, gadobutrol has been shown in a number of clinical studies to be safe while improving image enhancement with increased conspicuity and sensitivity over other commonly used GBCAs.

Keywords

Gadolinium-based contrast agent (GBCA), contrast-enhanced magnetic resonance imaging, stability, relaxivity, concentration

2

Article

Since the original conception and introduction of magnetic resonance imaging (MRI) gadolinium-based contrast agents (GBCAs) in the 1980s, many such contrast agents have been approved for use in contrast-enhanced MRI. These contrast agents have been used extensively in a large range of indications, particularly central nervous system examinations, as well as neurodegenerative diseases and tumoural disease processes, among others.


Since the original conception and introduction of magnetic resonance imaging (MRI) gadolinium-based contrast agents (GBCAs) in the 1980s, many such contrast agents have been approved for use in contrast-enhanced MRI. These contrast agents have been used extensively in a large range of indications, particularly central nervous system examinations, as well as neurodegenerative diseases and tumoural disease processes, among others.
Contrast enhancement has enabled improved tissue contrast and lesion characterisation and more sensitive detection of even very small lesions. The differences between the molecular structures of the gadolinium chelates affect their physicochemical characteristics. Greater importance has recently been placed on GBCAs and chelate stability, particularly with the possible association between the development of nephrogenic systemic fibrosis (NSF) in patients with kidney failure.1–3
In response to the growing interest in this topic, this article will address the differences between gadolinium-based MRI contrast agents with regard to their molecular structure and physicochemical properties, providing evidence of the advantages of some compounds in terms of stability, relaxivity and concentration.

Molecular Structure and Physicochemical Characteristics

GBCAs consist of the active substance gadolinium and a chelating agent. They can be categorised by their chemical structures into linear and macrocyclic agents and further subdivided by their charge (ionic or non-ionic). While differences do exist among the gadolinium compounds in terms of various physicochemical characteristics, stability, concentration and relaxivity appear to have a major clinical impact.

Stability

The chelation of gadolinium to organic ligands promotes bio-distribution within the extracellular interstitial space while enhancing renal filtration.2 After intravenous administration, GBCAs distribute into accessible extracellular spaces, with a distribution half-life of approximately 10 minutes.4 GBCAs are typically excreted by the kidneys, with an elimination half-life of approximately 90 minutes in patients with normal renal function. Over 90% of the injected dose is eliminated renally within the first 24 hours.2,5 By comparison, patients with reduced kidney function encounter a prolonged elimination half-life of several hours, depending on the degree of renal impairment, with >80% of the administered dose excreted over the subsequent week.6 Notably, renal impairment does not affect the extracellular distribution half-life.7
The greatest concern regarding the stability of gadolinium chelates relates to the risk of transmetallation, where gadolinium has the potential to be exchanged between ligands or even be released as a free Gd3+ ion and retained within the body. Great emphasis has been placed on gadolinium chelate stability in vivo and patient safety in recent years, particularly given the possible association with the occurence of NSF.1–3
With the gadolinium ion tightly caged within the ligand cavity by virtue of a rigid and pre-organised ring (see Figure 1), the macrocyclic gadolinium chelates are generally more stable than their linear counterparts.8,9 In a recent investigation into the stability of GBCAs, the rates of dissociation of Gd3+ ions were determined by incubating the gadolinium compounds in human serum over 15 days at 1mmol/l under physiological conditions (pH 7.4, 37°C).9 The non-ionic linear gadolinium chelates were found to be the least stable, with the greatest propensity to release Gd3+ ions into native human serum. Non-ionic chelates were followed by ionic linear compounds and finally macrocyclic compounds (see Figure 2A and Table 1). Notably, the release of Gd3+ ions by the macrocyclic compounds gadoterate (Dotarem), gadoteridol (ProHance) and gadobutrol (Gadovist) was below the detectable limit of quantification throughout the 15-day incubation period.
The addition of 10mmol/l of inorganic phosphate to human serum mimics one of the many conditions of acute renal insufficiency, although the phosphate levels are approximately triple those of typical clinical conditions of patients on dialysis. Under these conditions, the initial rates of Gd3+ release from formulated versions of gadodiamide (Omniscan) and gadoversetamide (OptiMARK) increased by approximately 100-fold. These two contrast agents are similar in that both are linear, non-ionic agents and have the lowest thermodynamic stability constant among the GBCAs. After 15 days, gadodiamide and gadoversetamide had released about 37% of the total amount of Gd3+, which is roughly 1.8-fold compared with native serum (see Figure 2B). While the ionic linear compounds also exhibited a 12–30-fold accelerated rate of Gd3+ release in the presence of elevated phosphate levels, the total amount of Gd3+ released after 15 days was comparable to that of native serum. Notably, the elevated phosphate levels had no discernable effect on the release of Gd3+ ions from the macrocyclic contrast agents. The GBCAs can therefore be divided into three distinct stability classes: non-ionic linear agents, ionic linear agents and macrocyclic agents.

Relaxivity and Concentration

The relaxivity and concentration of any given contrast medium are properties that contribute to an agent’s efficacy via the shortening of relaxation times and the resulting enhancement. By definition, relaxivity values are concentration-normalised data. Being independent of gadolinium concentration, the effects of local concentrations of contrast are not accounted for. The relaxivity of different gadolinium chelates had previously been shown to decrease at increasing magnetic field strengths (see Figure 3).10 Significantly different dependencies of relaxivities on field strength exist between GBCAs. Gadobutrol (Gadovist), gadobenate dimeglumine (MultiHance) and the liver-specific gadoxetic acid (Primovist) consistently exhibited higher relaxivity relative to the other compounds across the different magnetic fields that are commonly used in clinical practice.
T1 shortening – the parameter that directly influences the available signal intensity in T1-weighted, contrast-enhanced MRI – is also affected by the concentration of the contrast medium. The local contrast medium concentration in turn depends on a number of intrinsic, physical and technical factors, including:
• the cardiovascular physiology of a particular individual;
• the region of interest;
• the injection rate (ml/s);
• the injected dose (mmol); and
• the concentration of the injected contrast agent.
In one study using an animal model, identical bolus injections of gadobutrol at two different concentrations resulted in a higher concentration in the common carotid artery by approximately 30% with the 1.0M solution compared with the 0.5M solution.11 Many other studies have since demonstrated a concentration-dependent effect on T1 shortening and imaging enhancement.12–16

Greater Enhancement and Improved Detection

Stability, concentration and relaxivity each have an obvious impact on the clinical performance of a gadolinium chelate. While stability can be linked to safety, relaxivity and concentration can directly affect the quality of the obtained image via the mode of action and shortening of relaxation times. Advantages in terms of concentration have previously been proposed; for example, the relatively high gadolinium concentration of gadobutrol at 1.0M potentially allows for a 50% reduction in injection volume compared with equimolar dosages of the standard gadolinium chelates at 0.5M. A smaller and more highly concentrated bolus volume may allow for increased intravascular concentration. This is on top of advantages in perfusion-weighted imaging and multiphasic ultra-fast imaging techniques, including magnetic resonance angiography (MRA) and time-resolved MRA.17 It is also possible that a better-defined bolus stemming from this improved bolus effect may also enhance the contrast-to-noise ratio (CNR) and therefore the imaging characteristics of first-pass MRI techniques.18
Tombach et al. had previously compared gadobutrol at concentrations of 0.5 and 1.0M in an intra-individually controlled study with magnetic resonance brain perfusion imaging.19 With the higher-concentration formulation of 1.0M, a significantly smaller bolus width was achieved at the half maximum signal intensity decrease. Moreover, the smaller mean peak time observed, higher contrast and higher signal-to-noise ratio (SNR) ultimately led to better brain perfusion images than those obtained with the 0.5M formulation.
In another study where 0.5M gadopentetate (Magnevist), 0.5 M gadobenate and 1.0M gadobutrol were compared in a pelvic 3D MRA setting, the quality of images obtained with either gadobutrol or gadobenate were similar in their superiority over those obtained by gadopentetate when administered at an equal dose and injection rate.20 Undiluted gadobutrol performed better than its diluted counterpart against gadobenate, although statistically significant results were not reached. The results still suggest that a highly concentrated and compact bolus of gadobutrol can confer substantial improvements in image quality, quantifiable through significantly higher SNRs and CNRs.
Several studies have evaluated the effects of gadolinium chelates on the shortening of relaxation times and contrast enhancement, knowing that there is an influence on these outcomes by both relaxivity and concentration. Data originating from a rat brain tumour model show that the application of gadobutrol at the same gadolinium dose per weight results in markedly higher SNR and CNR values than gadopentetate and gadoterate at three Tesla (3T).21 Superior enhancement was observed with the double-concentrated gadobutrol at each post-contrast time-point assessed.
More recent studies have compared the use of 0.5 and 1.0M contrast agents in the imaging of brain tumours. One study used exogenous contrast-based, T2*-weighted, gradient-recalled echo-planar imaging to determine intracranial perfusion characteristics in 11 patients (six with intra-axial and five with extra-axial tumours). Patients received either 5ml of 1.0M gadobutrol or 10ml of 0.5M gadopentetate.22 The investigators assessed the maximal signal change in the region of interest as the primary outcome measure for perfusion-weighted imaging. They found that at 3T, the higher concentration of 1.0M gadobutrol provided significantly better delineation between grey and white matter. It also better defined the presence of highly vascularised brain tumour tissue than the standard 0.5M strength gadopentetate.
Visualisation of brain metastases has also been reported to be significantly improved with gadobutrol over gadopentetate and the detected lesions were more conspicuous.16 In this study, equal gadolinium dosages of each contrast agent were injected in a fully randomised sequence into 27 patients with at least one cerebral metastasis, with 18 hours between examinations with contrast-enhanced MRI. A total of 67 metastatic lesions were detected using gadobutrol enhancement compared with 65 lesions detected with gadopentetate; in two cases lesions were observed only with gadobutrol (see Figure 4). A comparison of images on a qualitative level by two fully blinded and experienced neuroradiologists revealed improved lesion conspicuity with gadobutrol in 10/27 of cases (p=0.002; see Figure 5). The comparison also revealed equivalent conspicuity in 17/27 cases, with no reported improvements in conspicuity for gadopentetate over gadobutrol.
The results from these small studies suggest that highly concentrated GBCAs confer significant advantages in imaging enhancement.

Summary and Conclusions

The differences between the physicochemical properties of GBCAs have an impact on the stability of the chelate. In vitro experiments have shown that the macrocyclic compounds (gadoterate, gadoteridol and gadobutrol) are the most stable, with an undetectable release of Gd3+ ions under physiological conditions.
High relaxivity, correlating to improved enhancement, has also been associated with certain gadolinium chelates over others. Studies suggest that highly concentrated formulations may be advantageous, with reduced injection volumes and potentially beneficial bolus effects with a smaller bolus volume and increased intravascular concentration for dynamic examinations. Indeed, emerging data from several studies now show that the higher-molar formulation of gadobutrol appears to confer significant advantages in imaging with respect to conspicuity in lesion visualisation and sensitivity in lesion detection compared with other commonly used gadolinium chelates at the standard 0.5M. â– 

2

References

  1. Abraham JL, Thakral C, Tissue distribution and kinetics of gadolinium and nephrogenic systemic fibrosis, Eur J Radiol, 2008;66:200–207.
  2. Bellin MF, Van Der Molen AJ, Extracellular gadoliniumbased contrast media: an overview, Eur J Radiol, 2008;66:160–67.
  3. Morcos SK, Nephrogenic systemic fibrosis following the administration of extracellular gadolinium based contrast agents: is the stability of the contrast agent molecule an important factor in the pathogenesis of this condition?, Br J Radiol, 2007;80:73–6.
  4. Caravan P, Ellison JJ, McMurry TJ, et al., Gadolinium(III) chelates as MRI contrast agents: structure, dynamics, and applications, Chem Rev, 1999;99:2293–2352.
  5. Oksendal AN, Hals PA, Biodistribution and toxicity of MR imaging contrast media, J Magn Reson Imaging, 1993;3: 157–65.
  6. Joffe P, Thomsen HS, Meusel M, The pharmacokinetics of gadodiamide in patients with severe renal insufficiency treated conservatively or undergoing hemodialysis or continuously ambulatory peritoneal dialysis, Acad Radiol, 1998;5:491–502.
  7. Tombach B, Bremer C, Reimer P, et al., Pharmacokinetics of 1M gadobutrol in patients with chronic renal failure, Invest Radiol, 2000;35:35–40.
  8. Tweedle MF, Hagan JJ, Kumar K, et al., Reaction of gadolinium chelates with endogenously available ions, Magn Reson Imaging, 1991;9:409–15.
  9. Frenzel T, Lengsfeld P, Schirmer H, et al., Stability of gadolinium-based magnetic resonance imaging contrast agents in human serum at 37 degrees C, Invest Radiol, 2008;43:817–28.
  10. Rohrer M, Bauer H, Mintorovitch J, et al., Comparison of magnetic properties of MRI contrast media solutions at different magnetic field strengths, Invest Radiol, 2005;40: 715–24.
  11. Muhler A, Weinmann HJ, Arterial concentrations following intravenous bolus injection of gadobutrol: comparison of 0.5 and 1.0 molar stock solutions, Presented at the Proceedings of the Society of Magnetic Resonance in Medicine and the European Society of Magnetic Resonance in Medicine and Biology, 1995, Berkeley, California, Society of Magnetic Resonance in Medicine, abstract 1128.
  12. Shang Y, Zheng ZZ, Li X, Direct MR arthrography of the shoulder at 3 Tesla: optimization of gadolinium concentration, J Magn Reson Imaging, 2009;30:229–35.
  13. Masi JN, Newitt D, Sell CA, et al., Optimization of gadodiamide concentration for MR arthrography at 3 T, AJR Am J Roentgenol, 2005;184:1754–61.
  14. Hajek PC, Sartoris DJ, Neumann CH, et al., Potential contrast agents for MR arthrography: in vitro evaluation and practical observations, AJR Am J Roentgenol, 1987;149:97–104.
  15. Kopka L, Funke M, Fischer U, et al., MR arthrography of the shoulder with gadopentetate dimeglumine: influence of concentration, iodinated contrast material, and time on signal intensity, AJR Am J Roentgenol, 1994;163:621–3.
  16. Anzalone N, Gerevini S, Scotti R, et al., Detection of cerebral metastases on magnetic resonance imaging: intraindividual comparison of Gadobutrol with Gadopentetate Dimeglumine, Acta Radiol, 2009;50(8): 933–40.
  17. Tombach B, Heindel W, Value of 1.0-M gadolinium chelates: review of preclinical and clinical data on gadobutrol, Eur Radiol, 2002;12:1550–56.
  18. Thilmann O, Larsson EM, Bjorkman-Burtscher IM, et al., Comparison of contrast agents with high molarity and with weak protein binding in cerebral perfusion imaging at 3 T, J Magn Reson Imaging, 2005;22:597–604.
  19. Tombach B, Benner T, Reimer P, et al., Do highly concentrated gadolinium chelates improve MR brain perfusion imaging? Intraindividually controlled randomized crossover concentration comparison study of 0.5 versus 1.0 mol/L gadobutrol, Radiology, 2003;226:880–88.
  20. Herborn CU, Lauenstein TC, Ruehm SG, et al., Intraindividual comparison of gadopentetate dimeglumine, gadobenate dimeglumine, and gadobutrol for pelvic 3D magnetic resonance angiography, Invest Radiol, 2003;38: 27–33.
  21. Attenberger UI, Runge VM, Jackson CB, et al., Comparative evaluation of lesion enhancement using 1 M gadobutrol vs. 2 conventional gadolinium chelates, all at a dose of 0.1 mmol/kg, in a rat brain tumor model at 3 T, Invest Radiol, 2009;44:251–6.
  22. Giesel FL, Mehndiratta A, Risse F, et al., Intraindividual comparison between gadopentetate dimeglumine and gadobutrol for magnetic resonance perfusion in normal brain and intracranial tumors at 3 Tesla, Acta Radiol, 2009;50:521–30.
3

Article Information

Disclosure

Editorial assistance was provided by Touch Briefings and supported by Bayer Schering Pharma AG.

Correspondence

Nicoletta Anzalone, Department of Neuroradiology, Scientific Institute H.S. Raffaele, Via Olgettina 60, 20132, Milan, Italy. E: anzalone.nicoletta@hsr.it

Received

2009-11-30T00:00:00